Download Brochures here
Description
Assay methods: End Point, 2 End Point Kinetic, Fix time etc.
Principle: Photoelectric colorimetry
Light Source: Halogen-Tungsten lamp with long life
Photometry range: 0~4 Abs
Resolution: 0.0001
Abs Wavelength: 340nm\405nm\450nm\492nm\510nm\546nm\578nm\630nm\700nm\800nm
Throughput: 200 tests/hour
Reagent tray: 40 reagent positions (include1detergent position,1 dilution position) Bar Code option for Reagents
Sample tray: 45sample positions, including detergent, standard, QC, STAT positions. Bar Code options for Samples
Reaction tray: 60 reaction cuvettes Sample volume: 2~50ul, with 1ul increment
Reagent volume: R1:10~400uL, R2: 10~400uL,with 0.5uL increment Minimum
reaction volume: 150uL
Refrigeration : On board cooling for Reagents , Samples , Calibrators , Standards , Control with ON/OFF Switch. 24 hours non-stop cooling system to ensure reagent at 2-8 degrees
Maximum reaction time: 10 minutes
Water Consumption: 2L/hour under working status
Clean unit: 8-step auto-washing system with detergent
Calibration: Calibration reset, select best test point by reaction curve, no need second calibration Line/non-line;multi-standards assay
Control rules: Westgard multi-rule, Cumulative sum check, Twin plot 3 level controls
for each ifem, analyxing and printing QC analysis diagram
Temperature control: incubator 37± 0.1
Power supply: AC 220V, 50Hz
Ambient: Operating temperature: 10~30 degrees
Relative humidity: <85% Atmospheric pressure: 86~106kPa Dimension : 560 mm x 410 mm x 360 mm Weight : 30 Kgs Support LIS interface